6.1 Printed Report Examples
HA-8190V 6-9
Example: Diagnosis results in the Fast mode
Example: Diagnosis results in the Variant mode
HbA1c reproducibility test
Fast
10-06-2017 10:00
HbA1c(IFCC) HbF
1 33 0.8
2 33 0.8
3 33 0.7
4 33 0.8
5 34 0.8
:
:
10 33 0.8
Mean
33.1
0.79
R
1
0.1
S.D.
0.3
0.03
C.V.
0.9
Print date
Current measurement mode
Result obtained by each measurement
Mean value of measurement results
Difference between maximum and minimum measurement
results
Standard deviation
Coefficient of variation
HbA1c reproducibility test
Var.
10-06-2017 10:00
HbA1c(IFCC) HbF
1 33 0.8
2 33 0.8
3 33 0.7
4 33 0.8
5 34 0.8
:
:
10 33 0.8
Mean
33.1
0.79
R
1
0.1
S.D.
0.3
0.03
C.V.
0.9
Print date
Current measurement mode
Result obtained by each measurement
Mean value of measurement results
Difference between maximum and minimum measurement
results
Standard deviation
Coefficient of variation