2.4 Measurement Mode Selection (Variant or Fast)
HA-8190V 2-15
d Perform the control measurement.
• See “2.7. HbA1c Control Measurement” on page 2-33.
e Check the latest calibration date and coefficients.
f Perform HbA1c automatic calibration as needed.
• See “2.8. HbA1c Calibration” on page 2-38.
NOTE:
If measurement mode is likely to change during consecutive measurements, perform a control measurement in
each mode as needed.
REFERENCE:
If there is a problem with the calibration dates displayed on the screen (e.g., when HbA1c calibration was
performed on a different day for each mode or when a long period of time has passed since the last HbA1c
calibration), it is recommended to perform HbA1c calibration anew.