2.5 Normal Measurement
HA-8190V 2-17
Items required for normal measurement
Prepare sample containers, adapters and sample racks for the samples to measure, in the sequence indicated in the
following flow chart. The underlined items are not included with the instrument.
*1: Anemia rack
Use the anemia rack for whole blood samples from patients previously diagnosed as anemic.
Samples on the anemia rack are measured at a lower dilution ratio than samples on the normal racks.
However, A1c values may not accurately represent the mean blood sugar level over the past 2 or 3 months, if the
patient shows signs of anemia because blood corpuscles are short-lived. Therefore, base any diabetes
determination on a comprehensive diagnosis of blood sugar level and other factors.
*2: Sample containers for whole blood
Sample tube: Recommended. Samples are stirred, so stable measurement results can be obtained without
variations due to blood sedimentation.
Sample cup: When small volume of sample is available or when measuring few samples. Samples are not stirred.
*3: Barcode labels and sample tubes
Prepare empty sample tubes. Sample IDs are read from the barcode labels attached to the sample tubes.
REFERENCE:
If sample rack transportation is set to “loop transportation”, the start rack (START) is needed in addition to the
sample racks specified in the flow chart above.
When
reading
sample IDs
Barcode label
Barcode label,
Sample tube*3
Barcode label,
Sample tube*3
Sample rack Normal rack
Whole blood pair rack
(WBlood PAIR)
Hemolysis pair rack
(HEMOL. PAIR)
Sample tube with cap
(Recommended)*2
Sample tube with cap
(Recommended)*2
Port guide
(gray)
Port guide
(gray)
Sample
container
Diluent
Sample cup*2 Sample cup
About 15 mm
diameter
About 13 mm
diameter
Port guide
(gray)
None
Sample
Anemia (Using anemia rack)*1Anemia, non anemia
Whole blood
Hemolysis
DILUENT90
Barcode label
Anemia rack (ANEMIA)
About 15 mm
diameter
About 13 mm
diameter
Port guide
(gray)
None
Adapter
Whole blood
Sample tube with cap
(About 13 mm diameter)
Port guide
Normal rack Hemolysis pair rack
(HEMOL. PAIR, with port guides)
Sample tube
(for barcode
label)
Barcode label
Sample cup
(for sample)
a12569014872a
Example 1:
When preparing samples in sample tubes
(About 13 mm diameter)
Hemolysis
Example 2:
When using the internal barcode reader
HEMOL. PAIR