Azbil Corporation Safety
Model MGG10C/14C - MagneW FLEX+/PLUS+ Converter iii
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter CE
Conformity Supplement
CE CONFORMITY: This product is in conformity with the protection requirements of the
following European Council Directive: 2014/30/EU, the EMC Directive and
2014/35/EU, Low Voltage Directive. Conformity of this product with any
other “CE Mark” Directive(s) shall not be assumed.
EMC Directive/Standard PC Conformity Notes
ELECTROMAGNETIC COMPATIBILITY: 2014/30/EU,
EMC Directive
EMISSIONS: EN61326-1:2013, Gr.1 Class A, Electrical
equipment for measurement, control and laboratory use
EN 55011:2009/A1: 2010, Gr.1, Class A, Industrial Control
Equipment, Radiated electromagnetic disturbances 30MHz -
1000MHz,
30MHz-230MHz quasi-peak limit 40dB(uV/m) at
10m
230MHz-1000MHz quasi-peak limit 47dB(uV/m)
at 10m
1
IMMUNITY: EN61326-1:2013, Electrical equipment for
measurement, control and laboratory use, EN 61326-2-3:2013,
Particular requirements
PERFORMANCE: Unless otherwise noted, the
performance of this product, at the specied levels
of electromagnetic interference, is within the
specications for for “Performance Under Rated
Conditions.”
EN 61000-4-2: 2009, ESD, Electrostatic Discharge B
B
4 kV Contact
8kV Air
EN 61000-4-3:2006 + A2:2010, Radio-frequency
electromagnetic eld, amplitude modulated 80 -2700 MHz
A 1kHz, AM80%
10 V/m (80 -1000 MHz)
3V/m (1.0 -2.0 GHz)
1V/m (2.0 -2.7 GHz)
1
EN 61000-4-4:2012, Electrical Fast Transients/Burst B 1kV 1
EN 61000-4-5: 2006, Surge B 1kV 1
EN 61000-4-6:2009, Conducted Radio-frequency, 150 KHz -
80 MHz
A 3V 1
EN 61000-4-8: 2010, Power frequency magnetic eld A 30A/m
50Hz
1
EN 61000-4-11:2004, Voltage Dip/short interruptions B
C
C
C
0.5, 1cycle 0% (100%)
10/12 cycle 40% (60%)
25/30 cycle 70% (30%)
250/300 cycle 0% (100%)
NOTES:
PC = Performance Criteria
1. Twist pair cables required for all I/O interface circuits.
In case of remote model,two core double shield cable in metal conduit pipe is required for the input line in
connection with detector.
Performance Criteria: Immunity includes the tests and severity levels specied in EN 61326-1-2013 and EN 61326-2-
3-2013.
LV Directive Conformity
LOW VOLTAGE DIRECTIVE: 2014/35/EU EN 61010-1: 2010, Safety requirements for electrical equipment for
measurement, control and laboratory use Part 1: General requirements