FARO Edge
September 2012
92
Chapter 6: Maintenance and Troubleshooting
Look in the probe compensation log and compare the X, Y and Z values. Each
value should repeat to within the accuracy of the device.
N
OTE: If one person is able to repeat better than another person, review probe
compensation procedure. A “green light” does not guarantee a good probe
compensation. It simply indicates that the data collected was within
specification.
Mounting
Most accuracy and repeatability problems can be attributed to improper
mounting of the Edge. These issues are not always obvious. An apparently stable
mount can actually be moving more than the accuracy of the device, directly
impacting results and causing system performance degradation.