Chapter 10: Configuring the FARO Edge in CAM2 Measure
Edit Probe
Click the EDIT button in the COMPENSATION dialog box to modify the details of
the current probe. Change any settings and click the MODIFY button to
continue.
To create a new probe:
1 Enter a new Name for the probe.
2 Enter the diameter of the probe in the correct unit. Check the status bar at the
bottom of your screen for the current units.
3 Select the A
UXILIARY SWITCH check box, and enter a De-bounce Time if
necessary.
4 Select the G
UIDANCE check box to use the guidance feature in Hole
Compensation. See “Hole Method - Guidance” on page 182, and “Hole
Compensation Method” on page 185.
5 Click the CREATE NEW button.
The new probe is now current. You must compensate the new probe.
Probe Compensation Overview
Probe compensation is a localized process by which a measurement device is
optimized to perform measurements accurately.
To understand probe compensation, you must first understand the Edge's
reference system. The Edge is factory compensated from the base to the last joint
or axis, and the position of this joint is defined by the coordinate system which
originates at the base of the Edge. The last axis of the Edge has its own
coordinate system, and the location of the center of the ball probe is reported in
the probe's coordinate system. After the probe's coordinates are established, these
Figure 10-4 Modify Probe dialog box
08M52E00_FaroArm_Edge.book Page 180 Thursday, October 4, 2012 9:31 AM