Chapter 10: Configuring the FARO Edge in CAM2 Measure
CAUTION: The probe must be well-seated in the hole when digitizing all
compensation points. Even one or two poorly digitized points significantly
affects the optimization process, which then has an effect on the accuracy of
the Edge.
The digitized compensation points then calculate and the C
OMPENSATION
S
TATUS updates. If the probe passes, the current date and time is added to the
probe information.
You can view the details for all of the compensations of a probe by clicking the
View Log button. See “View Log” on page 189.
• Digitize points in the hole and sweep to
position #3.
• Digitize points in the hole and rotate the handle
to position #4.
Figure 10-6 Hole Compensation Method