GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 8 Replacement Procedures 8-23
12.)The power supply can now be removed and set on the floor while the last power cable is accessed.
13.)Disconnect the motherboard power cable, as shown in Figure 8-17 on page 8-23 , by pushing down
on the back of the locking key in the center of the connector and pulling the connector out.
The power supply is now removed from the BEP. Set it aside for disposal.
Figure 8-16 Four Power Supply Screws in the Back
Figure 8-17 Disconnect Motherboard Power Cable
Motherboard Connector
Locking Key