GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 8 Replacement Procedures 8-71
8-14-1 Applications Software Upgrade Procedure (cont’d)
7.) As indicated, press any key to continue. The process is completely automated and provides
you with messages as to its status. R2.1.0 Software provides the message in Figure 8-79.
At this point you just have to wait while the software loads (Approximately 10 minutes).
When the process has completed, the window will turn dark red and you will see the following message
in Figure 8-80.
8.) When it completes loading the system will reboot.
9.) Remove the CD from the drive
If possible, while the system is rebooting remove the CD from the drive. If you didn’t do that don’t worry
you will get a dialog box like the one shown in Figure 8-81.
10.)REMOVE the Applications CD
11.)From the Start Applications dialog box, select START
.
12.)Select CANCEL
.
Figure 8-79 R2.1.0 Software Message
Figure 8-80 Software Load Complete
Figure 8-81 Start Software