GPT-9500 Series User Manual
48
Set the test time
4. Press the SAVE soft-key to complete.
CONT. Test Time
When it is either ACW, DCW or IR test, the TEST
TIME can be set CONT., which means the test time
will last infinitely until FAIL judgment occurs.
Identical with the regular setting for
TEST TIME, use the scroll wheel to
set CONT. for TEST TIME value.
Test time is set CONT.
Note
When setting greater than 40VA for DCW and
greater than100VA for ACW, respectively, the
maximum test time is 600 seconds followed by the
identical rest time.