GPT-9500 Series User Manual
68
5. Press the START button to perform
the OFFSET function. The
resistance of the tester, after the
OFFSET has finished, will be
added into the OFFEST field with
an icon in proximity shown below.
OFFSET reference is set
OFFSET CLEAR
It is available to clear the set
OFFSET reference value by
clicking the OF CLEAR soft-key.
The OFFSET icon will be disappeared meaning no
offset reference is set and the OF CLEAR soft-key
will be grayed out accordingly.
OFFSET reference is cleared
OF CLEAR soft-key is grayed out