10. System Trace Noise (Only for Analyzers without
Option 006)
Frequency Range Ratio System Trace Noise System Trace Noise
(Magnitude1
)
(Ph-3
30kHzto3GHz
A/R
<0.006
dB
rms
<O.O38O
rms
30
kHz
to 3
GHz
B/R
<0.006
dB
rms
<o.03s” rms
1 At
+5
dBm
into
test port, 3
kHz
IF bandwidth, and CW sweep.
Required Equipment for
508
Analyzers
Cable,
APC-7,
24-inch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP P/N 8120-4779
Required Equipment for
750
Analyzers
Cable,
75Q,
Type-N
24-inch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP P/N 8120-2408
Analgzer
warmup time: 1 hour
Perform this test to measure the system trace noise at a designated frequency in
both the A/R and
B/R
ratioed
measurements.
1. Connect the equipment as shown in Figure 2-31.
HP
8753E
NETWORK ANALYZER
PORT 1
U
PORT 2
CABLE
24 INCH
Figure 2-31. System Trace Noise Ikst Setup
System
Veriication
and
2-65
Performance
Teste