‘Jhble
7-5. VC0 Range Check Frequencies
Imtmment
Setting
Counter
Qeadng
1
31Bmz
30f0.030
MHZ
6o.oooooo
MHZ
60*0.060
MHZ
4. Check the counter reading at the frequencies indicated.
q
If the readings are within the limits specified, the probability is greater
than 90% that the fractional-N assemblies are functional. Either skip
ahead to the “A7
PuIse
Generator Check” or perform the more conclusive
“Al4
VCO Range Check with Oscilloscope” described below.
q
If the readings
fail
the specified limits, perform the
“Al4
VCO Exercise”.
Al4
VCO Range Check with Oscilloscope
1. Remove the
W9
HI OUT cable (A14Jl to A7) from the A7 assembly and
connect it to an oscilloscope set for 50 ohm input impedance. Switch on the
analyzer.
Menus and Error Messages”, for more information-on the F’RACN TUNE
mode.
3. Vary the fractional-N VCO frequency with the front panel knob and check
the signal with the oscilloscope. The waveform should resemble
F’igure
7-17,
Figure 7-18, and
F’igure
7-19.
If the fractional-N output signals are correct, continue source troubleshooting
by skipping ahead to
“A7
Pulse Generator Check”.
Source Troubleshooting
7-25