HP 8753E - Equipment Setup for Source Power Check

HP 8753E
612 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HP
8753E
NETWORK
ANALYZER
/
\
HP
438A
POWER
METER
ADAPTER
WC-7 to
N(f)
li
_-
.-,
:
I
HP
8481A
HP
8482A
POWER
5
POWER
SENSOR
SENSOR
s9648e
Figure 4-5. Equipment Setup for Source Power Check
2. Zero and calibrate the power meter. Press
LPreset]
on the analyzer to initialize
the instrument.
3.
on
the
analyzer,
press
LMenu)
~~~~~
~
Lk/m)
to
output
a
cw
3oo
kHz
signal. The power meter should read approximately 0
deem.
4. Press
[161
m
to change the CW frequency to 16 MHz. The output power
should remain approximately 0
dBm
throughout the analyzer frequency
range. Repeat this step at 1 and 3
GHz.
(For Option 006 include an additional
check at 6
GHz.)
If any incorrect power levels are measured, refer to Chapter 7, “Source
Troubleshooting.
n
4.14
Start Troubleshooting Here