System Verification Tests
These tests apply
mainly
to system-level, error-corrected verification and
troubleshooting. Tests 27 to 31 are associated with the system verification
procedure, documented in the “System Verification and Performance Tests”
chapter. Tests 32 to 43
facilitate
examinin
g the calibration coefficient arrays
(error terms)
resuhing
from a measurement calibration; refer to Chapter 11,
“Error Terms,” for
details.
27
28
29
30
31
32-43
Sys
Ver Init.
Recalls
the
initiaiization
state for system verification
from an HP 8753E verification disk, in preparation for a measurement
calibration. It must be done before service internal tests 28, 29, 30, or
31 are performed.
Ver Dev 1.
Recalls
verification Iimits from disk for verification device
#1 in
aII
applicable S-parameter measurements. It performs
pass/fail
Iimit testing of the current measurement.
Ver Dev 2.
Same
as 28
above for device
#2.
Ver Dev 3. Same as 28
above for device
#3.
Ver Dev 4. Same as 28
above for device
#4.
CaI
Coef
1-12.
Copies error term data from a measurement calibration
array to display memory. A measurement calibration must be complete
and active. The
de9nition
of calibration arrays depends on the current
calibration type. After execution, the memory is automatically
displayed. Refer to Chapter 11, “Error Terms,” for details.
lo-12
Service Key Menus and Error Messages