15. Test Port Receiver Phase Compression
Specifications
CW Frequency lkst Port
Phase1
300
kHz
to 3
GHz
Port 1
16”
3
GHz
to 6
GHz2
Port
1
57.5”
300
kHz
to
3
GHz
Port
2
16”
3
GHz
to 6
GHz2
Port 2
<7.5O
1 With 10 Hz IF bandwidth.
2 Only for
analyzer
with Option 006
-
30
kHz
tQ6GHZrsnge.
Required Equipment for
500
Analyzers
Cable,
APC-7,
24-inch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP P/N 8120-4779
Required Equipment for 75 ohm Analyzers
Cable, 75 ohm, Type-N,
24inch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 8120-2408
Anulgzer
warmup
time:
1
hour
Perform this test to verify the compression/expansion phase relationships of the
analyzer’s test port receiver samplers.
2.82
System Verification and
Performance Tests