13.
Press
.~&+$&@#‘$
to start the test. This part of the test will take about seven
minutes.
n If the analyzer displays Sampler Cor
-
FAIL, check the following:
a. The HP-IB address of your power meter is set at 13. Then rerun this
routine (“Sampler Correction Constants Routine”).
b. The HP
8482A
power sensor is connected to Port 1. Rerun this routine
(‘Sampler Correction Constants Routine
“).
19.
For Option 006 Instruments Only:
When the analyzer displays CONNECT
6
GHz
SENSOR B TO PORT 1, make the connections as shown in Figure 3-6.
Then press
@WJ~.
This part of the test will take about 20 seconds.
.._
.._..................-..-..
HP
8753E
NETWORK
ANALYZER
WITH OPTION
006
POWER
SENSOR
Figure 3-6. Connections for Sampler Correction at 6
GHz
20. When the analyzer displays CONNECT
<3
GHz
SENSOR A TO PORT 2, make the
connections as shown in Figure 3-7.
Adjustments and Correction Constants 3-23