Calculate the attenuation error values for this column by subtracting the
values
in column “C” from the values in column
“B”
(B
-
C = D).
4. The values in column “F” result from correcting the expected measurement
value by the amount of attenuator error.
Calculate the values in this column by subtracting the values in column “D”
from the values in column
“E”
(E
-
D =
F).
5. Transfer the values from column “F” in
‘Iable
2-l to column “F” in the
“Performance
Test
Record” for both test ports.
Power
Meter
Calibration
6. Zero and calibrate the power meter. (Refer to the power meter manual for
details on this procedure.)
7. Connect the equipment as shown in Figure 2-37.
2-60
System Verification and
Performance Tests