Error Messages
This section contains an alphabetical list of the error messages that pertain to
servicing the analyzer. The information in the list includes explanations of the
displayed messages and suggestion to help solve the problem.
Note
The error messages that pertain to measurement applications
are included in the HP 8753E Network Anulgzer User’s Guide.
BATTERY FAILED. STATE MEMORY CLEARED
Error Number The battery protection of the non-volatile SRAM memory has
183
failed. The SRAM memory has been cleared. Refer to the
“Assembly Replacement and Post-Repair Procedures” chapter
for battery replacement instructions See the “Preset State and
Memory Allocation,
n
chapter in the HP 87533 Network
Anulgm
User’s
Guide
for more information about the SRAM memory.
BATTERY LOW! STORE SAVE REGS TO DISK
Error Number The battery protection of the non-volatile SRAM memory is in
184
danger of failing. If this occurs, all of the instrument state
registers stored in SRAM memory will be lost. Save these states
to a disk and refer to the “Assembly Replacement and
Post-Repair Procedures” chapter for battery replacement
instructions. See the “Preset State and Memory Allocation,”
chapter in the HP 8753E Network
Anulgzer
User’s
Guide
for
more information about the SRAM memory.
CALIBRATION ABORTED
Error Number You have changed the active channel during a calibration so the
74
calibration in progress was terminated. Make sure the
appropriate channel is active and restart the calibration.
1 O-50