Backlight Intensity Check
Required Equipment and
Ytbols
Photometer. . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tektronix
J16
Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.‘Pektronix
J6503
Light
Occluder
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tektronix 016-0305-00
Antistatic Wrist Strap
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 9300-1367
Antistatic Wrist Strap Cord
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 9300-0980
Static-control Table Mat and Earth Ground Wire
. . . . . . . . . . . . HP P/N 9300-0797
Analgm-
warmup time:
30 minutes.
Photomder
warrn-up
time:
30 minutes.
Note
This procedure should be performed with a photometer and only
by qualified personnel.
1.
Press
~~)~~~~.~~~~~'.~~~~~~~~.
~~~~~~~Iloo]Lxl],
to
set
the
&play
-
. . . . .
~::::::.
..-.-.....
--
-......-......-.........
--
..-.
..-- --
--.-
-..
intensity at 100%.
,.,,,,,......,._.
_
.,.,.,.,.,
_
__
. . . . . . . . . . . .
. . . . . . . . . . .
_
,............,...,.........:....:::
/
,;.?..,
..,,
..;.
._
hy:.::*,7<;-
,,
..::::::::::::::..:..:<:<<~:‘~~;.
:
;.:
5
EEz;;
.::::::::::::::.::..
.:
2.
press
(77)
~~~~~~~,~
imm
L62)
Ixl)
~~~~.
~~~~~
to
.:
:.:
._...._
*
. . . . . . . . . . .
xi
. . . . .
.
;...:..A
.:.
.:.::..:
::.
>-.A
_
_
,.,.
__,/
set a white screen test pattern on the display.
3. Set the photometer probe to NORMAL. Press
I-)
on the photometer to
switch it on and allow 30 minutes of warm-up time. Zero the photometer
according to the manufacturer’s instructions
4. Center the photometer on the analyzer display as shown in
Figure
6-4.
6-8 Digital Control Troubleshooting