Frequency Accuracy Adjustment
Required Equipment and
‘Idols
Spectrum Analyzer
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP 8563E
RF Cable,
500
Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP P/N
8120-4781
Non-metallic Adjustment lb01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 8830-0024
Antistatic Wrist Strap . , . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP P/N 9300-1367
Antistatic Wrist Strap Cord
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 9300-0980
Static-control
‘lhble
Mat and Earth Ground Wire
. . . . . . . . . . . . HP P/N 9300-0797
Additional Required Equipment for
5OQ
Analyzers
Adapter
APC-7
to Type-N
(f)
. . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
11525A
Additional Required Equipment for
750
Analyzers
Minimum Loss Pad . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
11852B
Anul~m
warmup
time:
30 minutes.
This adjustment sets the VCXO (voltage controlled crystal oscillator) frequency
to maintain the instrument’s frequency accuracy.
1. Remove the upper-rear bumpers and analyzer top cover
2. Connect the equipment as shown in Figure 3-17.
348
Adjustments and Correction Constants