EasyManuals Logo

HP 8753E Service Manual

HP 8753E
612 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
Port 2 Noise Floor Level from 300
kHz
to 3
GHz
(IFm
=
3kHz)
i.
.;-...
.:‘<~$
18-
Press
@
8~~~~
@
Ck/m)
to change the IF bandwidth to 3
kHz.
20. When the analyzer
tlnishes
the sweep, notice the mean value.
21. Convert the measured linear magnitude mean value to log magnitude, using
this equation.
Power (dBm) = 20
*
[log&inear
magnitude mean
v&e)]
22. Write this calculated value on the “Performance Test Record.
n
23. This completes the
“Test
Port Input Noise Floor Level” procedure if your
analyzer does not have Option 006. Otherwise continue with the next
section.
Port 2 Noise Floor Level from 3
GHz
to 6
GHz
(IF BW = 3
kHz)
24.
Pressm@m[Stop-l@&J.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . .
25.
Ress
m
~~~~~~~~,
~~~~~.
. . . . . . . .
s...i..i
. . . . . . . . . . . .
4.s.....
. . . . . . . . . .
i,i__i..i
.,......................
::..
,.....,.,............................~...........~..~.~.
26. When the analyzer
Rnishes
the sweep, notice the mean value.
27. Convert the measured linear magnitude mean value to log magnitude, using
this equation.
Power (dBm) = 20
*
[log,,-,(Zineclr
magnitude mean value)]
28. Write this calculated value on the “Performance Test Record.”
240 System Verification and
Performance. Tests

Table of Contents

Other manuals for HP 8753E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8753E and is the answer not in the manual?

HP 8753E Specifications

General IconGeneral
BrandHP
Model8753E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals