Port 2 Noise Floor Level from 300
kHz
to 3
GHz
(IFm
=
3kHz)
i.
.;-...
.:‘<~$
18-
Press
@
8~~~~
@
Ck/m)
to change the IF bandwidth to 3
kHz.
20. When the analyzer
tlnishes
the sweep, notice the mean value.
21. Convert the measured linear magnitude mean value to log magnitude, using
this equation.
Power (dBm) = 20
*
[log&inear
magnitude mean
v&e)]
22. Write this calculated value on the “Performance Test Record.
n
23. This completes the
“Test
Port Input Noise Floor Level” procedure if your
analyzer does not have Option 006. Otherwise continue with the next
section.
Port 2 Noise Floor Level from 3
GHz
to 6
GHz
(IF BW = 3
kHz)
24.
Pressm@m[Stop-l@&J.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . .
25.
Ress
m
~~~~~~~~,
~~~~~.
. . . . . . . .
s...i..i
. . . . . . . . . . . .
4.s.....
. . . . . . . . . .
i,i__i..i
.,......................
::..
,.....,.,............................~...........~..~.~.
26. When the analyzer
Rnishes
the sweep, notice the mean value.
27. Convert the measured linear magnitude mean value to log magnitude, using
this equation.
Power (dBm) = 20
*
[log,,-,(Zineclr
magnitude mean value)]
28. Write this calculated value on the “Performance Test Record.”
240 System Verification and
Performance. Tests