Measure
Test
Port
1
Magnitude
Dynamic
Accuracy
33. Set the step attenuator to 10
dB.
34. To set up the dynamic accuracy measurement, press the following:
35.
W& for the
sweep
to
finish,
then press
(m)
f&$3
,;&
w,.
&&$g&,&&
*
....
.
...
. . . . .
...
.
/i...
~.~.ii~~...~.~.~
..__
i..::::..~~:
.
. . . . . .. . . . .
...
. . . . . . . . .
..
../...........
/.~.zu~
i....
LSS..
36. Set the step attenuator to 0
dB.
38. Write the mean value (which appears on the analyzer’s display) in the “Test
Port Measurement”
colrmm
of the “Performance Test Record.” This column
is also labeled “G.”
39. Repeat steps 36 through 38 for each setting of the step attenuator.
40. Calculate dynamic accuracy for each step by using the formula
IG
-
FI.
Place these values in the appropriate column of the “Performance Test
Record.
n
In
Case
of Difkulty
1. If the analyzer fails the test at ALL power levels, be sure you followed the
recommended attenuator settings as listed in the “Performance Test Record.”
Repeat this performance test.
2. If both test port measured values are out of specifications:
a. Recalibrate the power meter.
b. Repeat this performance test.
System
Verification and
2-87
Performance
Teete

Table of Contents

Other manuals for HP 8753E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8753E and is the answer not in the manual?

HP 8753E Specifications

General IconGeneral
BrandHP
Model8753E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals