Directivity
(EDF
and
EDR)
Description
Directivity is a measure of any detected power that is reflected when a load
is attached to the test port. These are the uncorrected forward and reverse
directivity error terms of the system. The directivity error of the test port
is determined by measuring the reflection
(Sll,
S22)
of the load during the
error-correction procedure.
S-cant
System Components
w
load used in the error-correction (calibration)
n
test port connectors
n
test port cables
Affected Measurements
Low
reflection device measurements are most affected by directivity errors.
HI
d
i i i i
i
i
i
i
i
I
(4
04
sg632s
Figure 11-2. Typical
EDF/EDR
without
and
with Cables
Error Terms
1
l-1
1

Table of Contents