Accessories
If the analyzer has passed all of the previous checks but is still making incorrect
measurements, suspect the system accessories Accessories such as RF or
interconnect cables, calibration or verification kit devices, limiters, and adapters
can
ail
induce system problems
Reconfigure the system as it is normally used and reconllrm the problem.
Continue
with
Chapter 9, “Accessories Troubleshooting.
n
Accessories Error Messages
n
POWERPROBESHUTDOWN!
The biasing supplies to a front panel powered device
(like
a probe
or millimeter module) are shut down due to excessive current draw.
Troubleshoot the device.
4-l
8 Start Troubleshooting Here