10. Observe the analyzer for the results of this
adjustment:
n If the marker value is less than 7, you have completed this procedure.
n If the marker value is greater than 7, readjust
‘TN
VCO ADJ” to 7.
Then perform steps 2 to 10 to con&m that the channel 1 and channel 2
markers are
stilI
above and below the reference line respectively.
w
If you cannot adjust the analyzer correctly, replace the
Al4
board
assembly.
Adjustments and Correction Constants
347