14.
T&t
Port Receiver Magnitude Compression
Specifications
Frequency Range
‘Jkst
Port
Magnitude1
300
kHz
to 3
GHz
Port 1
so.45
dB
3
GHz
to 6
GHz*
Port 1
10.80
dH
300
kHz
to 3
GHz
Port 2
LO.45
dB
3
GHz
to 6
GHz*
Port 2
~0.80
dB
1 With a 10 Hz IF bandwidth.
2
OnIy
for
maIyzers
with Option 006
-
30
IcHz
to
6
GHz
range.
Required Equipment for
5On
Analyzers
Cable,
APC-7,
24-inch
. . . . . . . . . . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
HP P/N 8120-4779
Required Equipment for 75 ohm Analyzers
Cable,
750,
Type-N,
24-inch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP P/N 8120-2408
Anul~zerwarmuptime:1hour
Perform this test to verify the compression/expansion magnitude levels of the
analyzer’s test port receiver samplers.
System Verification and 2-88
Performance Tests

Table of Contents