17.
‘I&t
Port Output/Input Harmonics (Option 002
Analyzers with Option 006 Only)
Specifications
T&t
Port Harmonic
Limit
<DUt
2nd
c-25
dBc
@I
+ 10
dBm
I
output
I
3rd
I<-25dEk@
+lOdBmI
I
Input Port 1
I
2nd
I<-15dEk@
+8dESm
I
I
Input Port 1
I
3rd
I<-3OdBc@
+8dESm
I
I
InDut
Port 2 I
2nd
1
<-15dBc@
+8dESm
1
I
InDut
Port
2
I
3rd
I
<-3OdEk@
+8dBm
I
Equipment Required
Cable,
APC-7,
24-inch
. . . . . . . . . . . . . . . . . . . . . . . . . . . , . . . . . . . . . . . . HP P/N 8120-4779
Attenuator
(2),
20
dB
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
8492A
Opt 020
Analggm
warmup time: 30 minutes
Perform this test to determine the spectral purity of the HP 8753E input and
output test ports
Note
The test port input 3rd harmonic specifications are
better
than
the test port output 3rd harmonic specifications.
System Verification and
2-101
Performance Tests