6. Observe the analyzer for the results of the routine:
n
If the analyzer displays
the
message
Serial Cor
DONE, you have
completed this procedure.
l
If the analyzer does not display DONE, then either the serial number that
you entered in steps 3 and 4 did not match the required format or a serial
number was already stored. Check the serial number recognized by the
analyzer:
b. Look for the serial number displayed on the analyzer screen.
c. Rerun this adjustment test.
n If the analyzer continues to fail this adjustment routine, contact your
nearest HP sales or service office.
Adjustments and Correction Constants
3-35