If the trace shows no improvement when the sampler correction constants are
toggled from off to on, perform the “Sampler Magnitude and Phase Correction
Constants (Test 53)” adjustment described in Chapter 3, “Adjustments and
Correction Constants” If the trace remains bad after this adjustment, the
A10
assembly is defective.
(4
04
sg643d
Figure 8-7. Typical Trace with Sampler Correction On and
off
Receiver Troubleshooting 8-13