NO FILE(S) FOUND
ONDISK
Error Number No files of the type created by an analyzer store operation were
45
found on the disk. If you requested a specific
hle
title, that
hle
was not found on the disk.
NO IF FOUND: CHECK R INPUT LEVEL
Error Number The first IF signal was not detected during pretune. Check the
5
front panel R channel jumper. If there is no visible problem
with the jumper, refer to the “Source Troubleshooting” chapter.
NO PHASE LOCK: CHECK R INPUT LEVEL
Error Number The
tist
IF signal was detected at pretune, but phase lock could
7
not be acquired. Refer to the “Source Troubleshooting” chapter.
NO SPACE FOR NEW CAL. CLEAR REGISTERS
Error Number You cannot store a calibration set due to insufficient memory.
70
You can free more memory by clearing a saved instrument state
from an internal register (which may also delete an associated
calibration set, if all the
instrument
states using the calibration
kit have been deleted.) You can store the saved instrument state
and calibration set to a disk before clearing them. After deleting
the instrument states, press
w
to run the memory packer.
1 O-54
Service Key Menus and Error Messages