Figure 2-2. Connections for Measurement Calibration Standards
16. When the analyzer finishes measuring the standard, connect the “short” end
of the open/short combination to reference test port 1.
18. When the analyzer hnishes measuring the standard, connect the 50 ohm
termination (supplied in the calibration kit) to reference test port 1.
19.
Pres
~~~~~~~~
*
.... ......
.
20. When the analyzer 6nishes measuring the standard, connect the “open” end
of the open/short combination to reference test port 2.
21.
mess
~~~~~~~~~.
-
.._........"
..:.-:
~.~.~.~.~.~.~..-~..~.~..~:.~.~::~
22. When the analyzer hnishes measuring the standard, connect the “short” end
of the open/short combination to reference test port 2.
24. When the analyzer
Gnashes
measuring the standard, connect the 50 ohm
termination to reference test port 2.