A4,
A5,
A6
Samplers and
A7
Pulse Generator
Replacement
1. Check the connector pins on the motherboard before reinstallation.
2. Reverse the order of the removal procedure.
Note
n
When reconnecting semirigid cables, it is recommended that
the connections be torqued to 10 in-lb.
n
Be sure to route
W8
and
W9
as shown. No excess wire should
be hanging in the All and
Al4
board
slots.
Routing the wires
in this manner will reduce noise and crosstalk.
14-26
Assembly Replacement and Post-Repair Procedures