1.
2.
Zero and calibrate the power meter. For more information of how to perform
this task, refer to the power meter operating manual.
Connect the equipment as shown in Figure 2-10.
ADAPTER
APC-7
HP
8753E
NETWORK ANALYZER
HP
43BA
HP
8753E
NETWORK ANALYZER
WITH OPTION 075
POWER SENSOR
Figure 2-10. Source Output Power Accuracy
lkst
Setup
System Verification and
2-25
Performance Tests