HAPSITE ER Operating Manual
7 If the Select Method First box was unchecked, select the Browse button in
order to select a method file. (See Figure 11-7.)
Figure 11-7 Browse button
8 Click the Browse button under Data Files. Select the desired data file for
library template creation. (See Figure 11-8.)
Figure 11-8 Calibration Control Panel with Data File Selected
NOTE: It is recommended to use a high or mid range standard for calibration
library development. Standards with low concentrations may have
peaks too small to be detected with default Search Settings.
9 Select Build/Edit Template. (See Figure 11-9.)
Figure 11-9 Build/Edit Template
10 Select the units. (See Figure 11-10.)
NOTE: Step 10 will be automatically completed if the information was entered
in the Data File Information window when the sample was run. Refer to
Step 3.
Figure 11-10 Unit Selection