11 - 5
HAPSITE ER Operating Manual
11 Set the Concentration Reference to Global or Analyte. (See Figure 11-11.)
NOTE: Select Global for standards that contain analytes that have the same
concentration. This is most common with liquid standards diluted into
a liquid.
NOTE: Select Analyte for standards that contain analytes with different
concentrations. This is the most common with liquid standards that are
diluted into a gas.
Figure 11-11 Global/Analyte
12 If Global is selected, enter in the concentration of the standard. Step 10 will be
automatically completed if the information was entered in the Data File
Information window when the sample was run. Refer to Step 3. (See
Figure 11-12.)
Figure 11-12 Global
13 If Analyte is selected, enter the volume of standard used for the selected data
file into the field highlighted below. For example, if an analyte was run at the
concentrations of 5 ppb, 10 ppb and 20 ppb, the factor for the 5 ppb data file
would be 1. For the 10 ppb, 2 would be the factor, and for the 20 ppb file, 4
would be the factor. The concentration of the standard would be entered into
the concentration column in Step 28. (See Figure 11-13.)
Figure 11-13 Analyte