11 - 10
HAPSITE ER Operating Manual
NOTE: When adding more than one quant ion, the following message will be
displayed. Click Yes to allow for the recalibration to continue. (See
Figure 11-28.)
Figure 11-28 Recalibration
28 Enter the lowest concentration of each analyte into the Conc column if
Analyte has been selected. If Global is selected, this step can be skipped.
(See Figure 11-29.)
Figure 11-29 Entering the Analyte Concentration
29 The Extracted Mass Peaks can also be edited to delete mass fragments. To
delete unwanted mass fragments, highlight the field and press the delete key.
Unwanted mass fragments would be those with intensities below 15%, unless
the fragmentation pattern is not very distinct. (See Figure 11-30.)
Figure 11-30 Extracted Mass Peaks