11 - 12
HAPSITE ER Operating Manual
33 To calibrate the library, click Browse to select the desired data files. (See
Figure 11-34.)
Figure 11-34 Browse to Select Data File
34 Select Calibrate Library. (See Figure 11-35.)
Figure 11-35 Calibrate Library
35 Check all of the data files and click the Start button under Peak Search. (See
Figure 11-36.)
Figure 11-36 Start
NOTE: Additional calibration points can be added to the curve by following
Step 8 through Step 35. Click OK when the A calibration point exists
already will not be added message is displayed. (See Figure 11-37.)
Figure 11-37 Confirming Additional Calibration Points