HAPSITE ER Operating Manual
39 It is possible to delete points from the calibration curve. (See Figure 11-41.)
NOTE: Removal of points in the middle of a calibration curve is contrary to
established analytical standards. Points can be removed from the highest
and lowest level of the curve, but this will affect the calibration range.
40 Click any number in the Conc. (Ratio to IS) column. The corresponding point
will be overlaid with a pink X. (See Figure 11-41.)
Figure 11-41 Clicking in the Conc. (Ratio to IS) Column
41 Use the up and down arrows to select the outlying point. (See Figure 11-42.)
Figure 11-42 Selecting the Outlying Point