5- 14 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Error Messages
B2961A/B2962A Specific Error
Error 203 Not able to execute while instrument is measuring
Error 210 Operation is not completed
Operation is still in progress. Wait for operation complete.
Error 211 Cannot switch low sense terminal with output on
Output relay must be off to switch low sense terminal.
Error 212 Output relay must be on
Error 213 Output relay must be off
Output relay must be on; External filter setting
Output relay must be off; Output resistance setting
Error 214 Display must be enabled
Display is currently disabled. Set remote display on.
Error 215 Remote sensing must be on
Remote sensing must be on to perform the requested operation.
Error 290 Not able to recall state: it is empty
Error 291 State file size error
Error 292 State file corrupt
Error 301 Emergency; Overvoltage status detected; Channel n
Overvoltage status was detected in channel n. All channel output is changed to 0 V
and the all output switch is opened. Execute the *TST? command.
Error 302 Emergency; Overcurrent status(245 V) detected; Channel n
overcurrent status (245 V) was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 303 Emergency; Overcurrent status(35 V) detected; Channel n
overcurrent status (35 V) was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.