Keysight B2961A/B2962A SCPI Command Reference, Edition 6 2- 9
Subsystem Command Summary
Setting Source Output and Measurement
[:SOUR[c]]:ARB:CURR:SIN:FREQ
freq
[:SOUR[c]]:ARB:VOLT:SIN:FREQ
freq
[:SOUR[c]]:ARB:CURR:SIN:FREQ?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:SIN:FREQ?
[DEFault|MINimum|MAXimum]
Set the frequency of the sinusoidal
waveform output.
freq=MINimum|MAXimum|DEFault|
0.001 to 10000, in Hz.
0.001
[:SOUR[c]]:ARB:CURR:SIN:OFFS
offset
[:SOUR[c]]:ARB:VOLT:SIN:OFFS
offset
[:SOUR[c]]:ARB:CURR:SIN:OFFS?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:SIN:OFFS?
[DEFault|MINimum|MAXimum]
Set the offset level of the sinusoidal
waveform output.
offset=MINimum|MAXimum|DEFault|
minimum to maximum source value, in A
or V. See
“Source Output Ranges” on
page 2-36.
0
[:SOUR[c]]:ARB:CURR:SIN:PMAR:PHA
S
phase
[:SOUR[c]]:ARB:VOLT:SIN:PMAR:PHA
S
phase
[:SOUR[c]]:ARB:CURR:SIN:PMAR:PHA
S?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:SIN:PMAR:PHA
S?
[DEFault|MINimum|MAXimum]
Specifies the position of the phase marker
for the sinusoidal waveform output.
phase=MINimum|MAXimum|DEFault|0
to 360, in degree.
[:SOUR[c]]:ARB:CURR:SIN:PMAR:SIG
N
output{,output}
[:SOUR[c]]:ARB:VOLT:SIN:PMAR:SIG
N
output{,output}
[:SOUR[c]]:ARB:CURR:SIN:PMAR:SIG
N?
[:SOUR[c]]:ARB:VOLT:SIN:PMAR:SIG
N?
Selects the trigger output at the phase
marker of the sinusoidal waveform
output.
output=INT1|INT2|EXT1|EXT2|
EXT3|EXT4|EXT5|EXT6|EXT7|EXT8|
EXT9|EXT10|EXT11|EXT12|EXT13|
EXT14
EXT1
Command Summary Reset setting