2- 10 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Command Summary
Setting Source Output and Measurement
[:SOUR[c]]:ARB:CURR:SIN:PMAR:STA
T state
[:SOUR[c]]:ARB:VOLT:SIN:PMAR:STA
T state
[:SOUR[c]]:ARB:CURR:SIN:PMAR:STA
T?
[:SOUR[c]]:ARB:VOLT:SIN:PMAR:STA
T?
Enables or disables the trigger output at
the phase marker of the sinusoidal
waveform output.
state=1|ON|0|OFF
OFF
[:SOUR[c]]:ARB:CURR:SQU:END:TIME
time
[:SOUR[c]]:ARB:VOLT:SQU:END:TIME
time
[:SOUR[c]]:ARB:CURR:SQU:END:TIME
?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:SQU:END:TIME
?
[DEFault|MINimum|MAXimum]
Set the end time of the square waveform
output. Waveform period must not exceed
1000 seconds.
time=MINimum|MAXimum|DEFault|0 to
1000 seconds.
0
[:SOUR[c]]:ARB:CURR:SQU:STAR[:LE
V] level
[:SOUR[c]]:ARB:VOLT:SQU:STAR[:LE
V] level
[:SOUR[c]]:ARB:CURR:SQU:STAR[:LE
V]?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:SQU:STAR[:LE
V]?
[DEFault|MINimum|MAXimum]
Set the start level of the square waveform
output.
level=MINimum|MAXimum|DEFault|
minimum to maximum source value, in A
or V. See
“Source Output Ranges” on
page 2-36.
0
[:SOUR[c]]:ARB:CURR:SQU:STAR:TIM
E time
[:SOUR[c]]:ARB:VOLT:SQU:STAR:TIM
E time
[:SOUR[c]]:ARB:CURR:SQU:STAR:TIM
E?
[DEFault|MINimum|MAXimum]
[:SOUR[c]]:ARB:VOLT:SQU:STAR:TIM
E?
[DEFault|MINimum|MAXimum]
Set the start time of the square waveform
output. Waveform period must not exceed
1000 seconds.
time=MINimum|MAXimum|DEFault|0 to
1000 seconds.
0
Command Summary Reset setting