2 General Power Meter Functions
116 Keysight N1911A/1912A User’s Guide
Optional measurement configurations:
TRIG:DEL 1ms
TRIG:HOLD 10us
OUTP:TRIG ON
SENS:SWE1:OFFS:TIME 100us
SENS:SWE1:TIME 100us
INIT:CONT ON
–*OPC
*ESR?
FETC?
Average mode This feature is applicable for 8480 Series, N8480 Series,
E-Series E4410, E-Series E9300, E-Series E9320 or P-Series N1920 sensor when
connected. Below are examples of SCPI commands when used in average mode.
Send the SCPI commands below to meter using remote interface.
SENS:DET:FUNC AVER
TRIG:SOUR EXT
SENS:FREQ:START 50 MHz
SENS:FREQ:STOP 4 GHz
SENS:FREQ:STEP 2048
Optional measurement configurations:
OUTP:TRIG ON
SENS:MRAT DOUB
SENS:AVER:COUN 128
INIT:CONT ON
–*OPC
*ESR?
FETC?