Keysight N9038A MXE Service Guide 631
Assembly Replacement Procedures
Front Frame Assembly
Figure 18-75 LED and Display Bracket Separation
Replacement
1. Refer to Figure 18-74. Place the display bracket into position in the Front
Frame Assembly. Reconnect the flex circuit A1W1 (1) and A1W3 (2) to
the front panel interface board (4).
2. Replace the screw (3) (0515-0372) that secures the display bracket to the
Front Frame Assembly. Torque to 9 inch pounds.
3. Replace the front panel shield.