X = Must Perform in the order listed
* = Must Perform if calibration file cannot be transferred to new disk drive
b
See the "Additional Tasks" section in this chapter for detailed instructions on these tasks
c
To set the instrument time and date select Start, Control Panel, Date Time Language and Regional Options, Date and Time
d
Before replacing the A5 Disk Drive, see the “Calibration Data Backup and Restore” section in this chapter for details
e
See Chapter 19, “Instrument Software” for instructions on updating the instrument software to the latest version
f
Only if option CR3 is installed
g
Found in instrument Service Menu under System, Service, RF Preselector, Characterize RF Preselector, All Bands. See Chapter 16 - Service & Diagnostics Menu for details.
i
Not applicable for option 544 instruments
h
Only if option TDS is installed
Additional Tasks
b
Utilities
a
Adjustments
a
Performance Verification
a
a
Calibration Software required for all performance tests, adjustments and utilities. For details see http://www.keysight.com/find/calibrationsoftware