Keysight N9038A MXE Service Guide 65
Boot Up and Initialization
Potential Problems During Boot Process
6. Verify that the on screen of the 4.8 GHz Calibrator signal is 28 dBm
(+/-2 dB), as shown in Figure 2-11.
Figure 2-11 4.8 GHz Calibrator Signal
7. If the signal level is not correct refer to Chapter 5, “RF Preselector
Section.” Chapter 6, “RF Downconverter Section. and Chapter 15, “Block
Diagrams.” for further troubleshooting information.