712                                       Keysight N9038A MXE Service Guide
Tests
Quasi-Peak Detector Accuracy
Error = Measured - (Power Meter + 106.99)
13.Set the EMI receiver input attenuator to 20 dB by pressing AMPTD Y Scale, 
Attenuation, 20 dB.
14.Setup the IGUU 2916 EMI pulse generator as follows:
a. Main Generator: Band A
b. Pulse Frequency: 25 Hz
c. Amplitude: 60 dBμV
15.Connect the IGUU 2916 EMI pulse generator Output Main Generator Band 
A/B to the EMI receiver RF Input.
16.Put the EMI receiver in single sweep operation by pressing Single.
17.Take a single sweep on the EMI receiver by pressing Restart.
18.Once the sweep has completed find the peak of the detector output by 
pressing Peak Search.
19.Record the marker value in Table 22-28 as the Measured value for the 
Band A reference PRF.
20.Calculate the Error with the following formula and enter the value in Table 
22-28.
Error = Measured - 60 dBμV - Error (Table 22-27)
21.Put the marker in Delta mode by pressing Marker, Delta.
22.Turn on the continuous peak function by pressing Peak Search, More, 
Continuous Peak Search On.
23.Change the Pulse Frequency on the IGUU 2916 EMI pulse generator to 
100 Hz.
Table 22-27  Quasi-Peak Sine-Wave Voltage Accuracy Band A - CISPR 16-1-1: 2010 Section 4.3
Sine Wave 
Frequency 
(kHz)
Band A
Power Meter 
(dBm)
Measured (dBμV) Lower Limit 
(dB)
Error 
(dB)
Upper Limit 
(dB)
100 −2.5 +2.5
Table 22-28  Quasi-Peak Absolute Calibration Band A - CISPR 16-1-1: 2010 
Section 4.4.1
Pulse 
Frequency 
(Hz)
Band A
Measured (dBμV) Lower Limit 
(dB)
Error 
(dB)
Upper Limit 
(dB)
25 −2.0 +2.0