732 Keysight N9038A MXE Service Guide
Tests
Peak Detector Accuracy
9. Change the detector used by the EMI receiver to Peak by pressing Trace /
Detector, More, Detector, Peak.
10.Put the EMI receiver in single sweep operation by pressing Single.
11.Turn on the continuous peak function by pressing Peak Search, More,
Continuous Peak Search On.
12.Take a single sweep on the EMI receiver by pressing Restart.
13.Once the sweep on the EMI receiver has completed and the auto peak
value is found record the marker amplitude as the Measured value for the
pulse frequency being measured in Table 22-45.
14.Without changing any other setting on the EMI receiver or the IGUU 2916
EMI pulse generator change the Pulse Frequency on the pulse generator
and the Sweeptime of the EMI receiver to the other values listed in Table
22-45 and repeat steps 12 and 13 for each.
15.Calculate the Error for each of the Measured values in Table 22-45 with
the following formula and enter the value in Table 22-45.
Error = Measured – 40 dBµV – 12.61 dB
Table 22-45 Peak Response to Pulses Band D – CISPR 16-1-1 :2010 Section 5.4
Pulse
Frequency
(Hz)
Sweeptime
(sec)
Band D
Measured
(dBµV)
Lower Limit
(dB)
Error
(dB)
Upper Limit
(dB)
20000 0.001 -2.0 +2.0
10000 0.001 -2.0 +2.0
5000 0.002 -2.0 +2.0
1000 0.02 -2.0 +2.0
500 0.02 -2.0 +2.0
100 0.1 -2.0 +2.0
50 0.1 -2.0 +2.0
10 2.0 -2.0 +2.0
1 2.0 -2.0 +2.0