84 Keysight N9038A MXE Service Guide
Instrument Messages
Event Messages
–221 Settings conflict;
Destination trace for
Trace Math cannot be
a trace operand
The resulting trace data (from doing a trace math function) cannot be put into
the any of the traces that are being used by the math operation
–221 Settings conflict;
Downconv only
available when DUT is
Amplifier
SCPI only message. The System Downconverter can only be set to ‘On’ when
the DUT type is amplifier. Change DUT type to Amplifier if the System
Downconverter is required
–221 Settings conflict;
EDGE EVM only
supports EDGE TCH
burst type.
–221 Settings conflict;
Electronic attenuator
is disabled
You are using the mechanical attenuator, and have not enabled the electronic
attenuator. You cannot set the value of the electronic attenuator because it
automatically sets/changes when enabled
–221 Settings conflict;
Electronic attenuator
is not available
above 3.6 GHz
The maximum frequency of the electronic attenuator is 3.6 GHz. This is
because of switching capacitance
–221 Settings conflict;
Electronic attenuator
unavailable in
current state
–221 Settings conflict;
Electronic attenuator
unavailable with
Preamp on
The internal preamp is on. Electronic attenuator cannot be used while you are
using the internal preamp
–221 Settings conflict;
EMI Detectors are not
available in FFT
sweep
QPD, EMI Average, EMI Peak, and MIL Peak are not allowed when in the
manually selected FFT sweep mode
–221 Settings conflict;
External Mixer not
available
A command has been sent to reference the External Mixer in a model that
does not contain it
–221 Settings Conflict;
FAST method can only
be used while Radio
Std is W-CDMA
–221 Settings conflict;
Feature not available
in this View
Some functionality is available in one View, but not in another. (See the Views
under the
View/Display key.) This error occurs if you send a SCPI
command or push a grayed-out key that is not available in the current
selected View
Err# Message Verbose/Correction Information