Keysight N9038A MXE Service Guide 85
Instrument Messages
Event Messages
–221 Settings conflict;
Feature not supported
for selected source
You have selected a feature that the selected source does not support
–221 Settings conflict;
Feature not supported
for this Input.
Some functionality is not available when certain Inputs are selected. For
example, Trigger Holdoff is not available for the BBIQ input
–221 Settings conflict;
Feature not supported
for this measurement.
Some functionality is available in one measurement, but not in another. (See
the measurements under the
Meas key.) This error occurs if you send a
SCPI command or push a grayed-out key that is not available in the current
selected measurement
–221 Settings conflict;
Feature not supported
for this model number
This functionality is not a part of the instrument you are using, but may be
found in other models in the X-Series
–221 Settings conflict;
FFT IF Gain High not
available when Swept
IF Gain = Manual Low
When Swept IF Gain is manually set to Low, you cannot set the FFT IF Gain to
High because that would make the Reference Level couplings wrong in FFT
mode
–221 Settings conflict;
FFT method is
unavailable for level
gating
If you are using level gating, you cannot select the FFT Gate Method
–221 Settings conflict;
FFT sweep type is not
available while in
Gated LO
The gated LO function turns the LO on and off as it sweeps, so the FFT sweep
type is not available if you have selected gated LO
–221 Settings conflict;
FFT sweep type is not
available while in
Gated Video
The FFT sweep type is not available if you have selected the gated video
function
–221 Settings conflict;
FFT Sweeps
unavailable in
Tracking Source mode
Since FFTs do not sweep, you cannot use a Tracking Source while doing FFTs
–221 Settings conflict;
Fixed marker adjust
not available while
Marker Function is on
If a Marker Function is on for a Fixed marker, the marker's reported value is
derived from the function. Therefore, you cannot directly set the X or Y value
of a Fixed marker that has a marker function turned on
–221 Settings conflict;
Fixed Marker Y value
is not adjustable
with Normalize On
If Normalize is on the Amplitude scale is in dB units, so adjusting the Y value
of a Fixed marker is not possible